Microhardness testing providers with MicroVision Laboratories near me? The scan from left to right shows a high tin concentration (green line) while areas of higher lead concentrations (blue line) were not intersected by the line scan. At the interface between the tin/lead solder and copper (red line), there is a mixture of the solder and copper which is the intermetallic layer. The EDS Map provides a nice visual mixture of colors which shows the intermetallic layer while the line scan clearly shows the intermetallic with the elemental graph.

The SEM was used to examine the crystal morphology, and the EDS spectrum showed primarily carbon and oxygen, with small amounts of nitrogen and phosphorous. This indicated an organic material as the primary component. Because the SEM-EDS analysis showed the material was primarily a carbon based organic crystalline material, a Fourier transform infrared spectroscopy (FTIR) examination was performed on the suspect material. This analysis provides necessary information about the functional groups of the organic material in order to identify the unknown organic.

MicroVision Labs is owned and operated by a career microscopist, John Knowles, who understands the needs of our clients. Our emphasis on helping our clients solve problems, not just providing data, sets us apart from other labs. We have the technology and knowledge to find answers to your most difficult challenges, helping you succeed at every step. Can I come in to see my samples analyzed? Yes, our clients are always welcome to come in while their samples are being analyzed. For much of the work we do, it is mutually beneficial for our clients to be present to help direct their project since they can provide expertise about their samples. Some of the services we provide such as polished cross sections have time consuming steps making it impractical for a client to stay to watch everything. In those cases it is recommended that you come in initially to explain what you need done and come back at a later time to see the finished product. Discover extra details on see the microvision labs website.

An affected floor tile was submitted to determine if the previous mold testing had missed a source on the tile backing or mastic. Additionally, a new tile from the same manufacturing lot was submitted for comparison. The process of preparing and examining the sample and reference tile was documented. Areas with darkened surface features were imaged and then cut out and examined. While the dark spots looked very discrete when examined by eye, under top light polarized microscopy they appeared more diffuse at the outer edges. The darkest areas surrounded what appeared to be particles embedded in the surface.

Have you always been located in Chelmsford, MA? No, for the first four years MicroVision Labs operated at 15 A Street, Burlington, MA. In 2007 we moved to our present location in Chelmsford, MA. What business designation does MicroVision Labs have? MicroVision Labs is designated as a veteran owned small business. How many staff members does MicroVision Labs have? MicroVision Labs is a small business employing 5-10 full-time, part-time, and contract employees.

Energy Dispersive X-Ray Spectroscopy (EDS): While in a Scanning Electron Microscope (SEM), samples are exposed to high energy electrons in a vacuum, which generates X-rays through secondary electron transitions. Variations in electron configuration specific to each element generate different energy electrons, and thus different signature energy peaks, indicating which elements are present in the sample. Analysis is performed only on areas which are exposed to the electron beam, facilitating precise control of the analyzed area. This means the composition of very small areas or particles in a sample can be taken. Since EDS is performed in the SEM chamber, a quick and easy interrogation of the surface materials as viewed on the SEM is possible. This can be expanded to include the entire sample, please see our Elemental Mapping page. Additionally, relative amount of the elements present can be calculated, generating composition percentages. Explore extra details at here.